Haku

Experiences in digitizing and digitally measuring a paper-based ECG archive

QR-koodi

Experiences in digitizing and digitally measuring a paper-based ECG archive

Abstract

Background: No established method for digitizing and digital measuring of paper electrocardiograms (ECG) exists. We describe a paper ECG digitizing and digital measuring process, and report comparability to manual measurements.

Methods: A paper ECG was recorded from 7203 health survey participants in 1978–1980. With specific software, the ECGs were digitized (ECG Trace Tool), and measured digitally (EASE). A sub-sample of 100 ECGs was selected for manual measurements.

Results: The measurement methods showed good agreement. The mean global (EASE)-(manual) differences were 1.4 ms (95% CI 0.5–2.2) for PR interval, −1.0 ms (95% CI −1.5–[−0.5]) for QRS duration, and 11.6 ms (95% CI 10.5–12.7) for QT interval. The mean inter-method amplitude differences of RampV5, RampV6, SampV1, TampII and TampV5 ranged from −0.03 mV to 0.01 mV.

Conclusions: The presented paper-to-digital conversion and digital measurement process is an accurate and reliable method, enabling efficient storing and analysis of paper ECGs.

Tallennettuna: